Power Supply Units And Ups Modules
Single-phase power supply 24 V DC (20 A)
Single-phase power supply unit (20 A) for constant direct current supply to the nurse call system.
Part Number: 89954R6
Single-phase power supply unit (20 A) for constant direct current supply to the nurse call system. Compact construction using modern primary clocked circuit technology (according to EN 6100-3-2) suitable for top-hat rail mounting in sub-distribution. Using the assembly kit (safety cover) 89954 MC, wall mounting is also possible outside sub-distribution. The output of the power supply unit is idle-, overload- and short-circuit proof.
Features
Connectors output:
- Push-In, max 2.5 mm2 for input
- Push-In, max 6.0 mm2 for output
- earth leakage current:
Signalling DC-OK:
- Uout > 21,5 V DC (LED)
- Uout > 21,5 V (typ.) max. 30 V/1A (relay)
Recommended circuit breaker:
- 20 A (characteristic B/C)
Switch-on current limitation:
- < 30 A, NTC
Accessories
89954C8 - UPS module 24 V DC (20 A) for 20 A power supply unit.
Others
- Version
- for top-hat rail mounting in sub-distribution
- Protection low voltage
- SELV
- Operation temperature
- 0 °C - 60 °C at free convection cooling
- Storage Temperature
- -40 °C ... 85 °C
- Ambient temperature
- -25 °C ... 70 °C
- Type of protection
- IP20 (IEC 60529)
- Rated voltage (V)
- 230 V AC
- Nominal frequency
- 50 Hz hertz
- Output voltage
- 23 V DC ... 28,5 V DC (adjustable)
- Output current
- 24 V / 20 A, 28 V / 18 A ampere
- Air humidity
- < 20 %
- Weight (kg)
- approx. 1,7 kg kilogram
- Dimensions
- W: 95 mm H: 127 mm D: 169 mm
Others
- Version
- for top-hat rail mounting in sub-distribution
- Protection low voltage
- SELV
- Operation temperature
- 0 °C - 60 °C at free convection cooling
- Storage Temperature
- -40 °C ... 85 °C
- Ambient temperature
- -25 °C ... 70 °C
- Type of protection
- IP20 (IEC 60529)
- Rated voltage (V)
- 230 V AC
- Nominal frequency
- 50 Hz hertz
- Output voltage
- 23 V DC ... 28,5 V DC (adjustable)
- Output current
- 24 V / 20 A, 28 V / 18 A ampere
- Air humidity
- < 20 %
- Weight (kg)
- approx. 1,7 kg kilogram
- Dimensions
- W: 95 mm H: 127 mm D: 169 mm
- Version : for top-hat rail mounting in sub-distribution
- Protection low voltage : SELV
- Operation temperature : 0 °C - 60 °C at free convection cooling
- Storage Temperature : -40 °C ... 85 °C
- Ambient temperature : -25 °C ... 70 °C
- Type of protection : IP20 (IEC 60529)
- Rated voltage (V) : 230 V AC
- Nominal frequency : 50 Hz hertz
- Output voltage : 23 V DC ... 28,5 V DC (adjustable)
- Output current : 24 V / 20 A, 28 V / 18 A ampere
- Air humidity : < 20 %
- Weight (kg) : approx. 1,7 kg kilogram
- Dimensions : W: 95 mm H: 127 mm D: 169 mm